| 98 | 1 | 11 |
| 下载次数 | 被引频次 | 阅读次数 |
热工水力实验测量仪表需要高效准确地完成计量校准和量值溯源,而目前基于过程校验仪的原位计量校准存在较多影响因素,难以准确量化不确定度分量。可柔性配置的原位校准标准装置主要基于校准溯源性评定技术、总线兼容技术、柔性化及并行测试技术进行设计,可柔性适应多种规模和仪表类型的反应堆工程热工实验测试系统。该装置的综合不确定度满足大部测试仪表的量值溯源要求,可有效提升实验用测试仪表的校准效率,是实验室计量校准模式的有益补充。
Abstract:Thermal-hydraulic experiment instrument need to achieve measurement calibration and measurement traceability efficiently and accurately, the influence factors of in-situ calibration mode based on process calibrator are difficult to quantification. Flexible configured in-situ calibration equipment is designed based on calibration traceability assessment technique, bus compatibility technology, flexible and parallel test technology. The in-situ calibration equipment adapt to a variety of sizes and types thermal experiment apparatus instruments calibration.The synthesize uncertainty and well-adaptable of in-situ calibration system can meet the need of various type of reactor engineering experiment instrument. The in-situ calibration system is a beneficial complement to the laboratory calibration.
[1]方立德,徐潇潇,马鑫月,等.计量仪器的远程量值传递与溯源技术探讨[J].中国测试, 2021,47(3):1-8.
[2]罗林聪,邢鑫,朱君,等.工业现场蒸汽流量计在线计量校准测试技术研究及应用[J].中国测试, 2020,46(z1):89-94.
[3]郭荣斌,赵秀才.自动测试系统的发展趋势[J].国外电子测量技术, 2014,33(6):1-4.
[4]季近健,孟晨.通用ATS总体技术指标确定方法研究与实现[J].中国测试, 2011,37(2):80-83.
[5]张琳,容晓峰,陈耀明.基于多总线的自动校准/校准测试系统设计[J].电子科技,2009,22(11):53-56.
[6]陈超. LXI总线协议实现与LXI-GPIB接口转换器设计[D].哈尔滨:哈尔滨工业大学,2010.
[7]郭论平. LXI接口VISA设计[D].哈尔滨:哈尔滨工业大学,2012.
[8]刘正升,万程亮.自动测试系统中新技术的发展及应用[J].中国测试, 2009,35(4):58-61
[9]丁蔚,彭诚,李军,等.基于虚拟仪器面向对象编程技术的计量校准程序[J].计量学报,2019,40(S1):69-74.
[10]肖勇.通用自动测试系统软件平台中仪器互换技术的研究[J].计量与测试技术,2020,47(2):22-24.
[11]程勤,王剑昆,柴艳丽,等.并行校准技术在航空装备计量保障平台中的应用[J].计测技术,2012,32(3):64-67.
基本信息:
中图分类号:TL33
引用信息:
[1]胡润勇,黄钰,赵豆,等.热工水力试验测量仪表原位校准装置设计[J].中国测试,2022,48(S1):207-210.
2022-08-31
2022-08-31